Details
Intended Use
This product is mainly used for observation and localization of superficial subcutaneous vessels in venipuncture departments and varicose vein operating rooms, assisting medical staff in venipuncture procedures such as venipuncture examination, infusion, and blood drawing.
Intended Use Environment
Intended use environment: This product is intended for use in medical institutions and clinics, with an ambient temperature of 0℃~40℃, relative humidity of 10%~80%, power supply of 5Vdc 2A, and atmospheric pressure of 86~106KPa.
Applicable Population
This product is suitable for all patients requiring venipuncture.
Applicable Sites
All sites with veins;
Contraindications
Patients allergic to 850nm infrared light are prohibited from using this product.
Structure and Composition
This product is composed of a main board, an optical engine, and an optical bracket.
Product Specifications
Safety Classification | |
|---|---|
| Electric Shock Protection Type | Internal battery powered equipment |
| Electric Shock Protection Class | No application part |
| IP Protection Class | IPX0 |
| Operation Mode | Continuous operation |
| Overvoltage Category | No |
| Risk Classification | Non-exempt |
Specification Parameters | |
| Output Light Wave Range | 850nm |
| Detection Depth | 1-15mm |
| Optimal Working Distance | 150mm-250mm |
Power Supply | |
| Power Input Range | 5Vdc 2A |
| Input Power | 10VA |
Operating Environment Conditions
(1) Temperature: 0℃~40℃
(2) Relative humidity: 10%~80%, non-condensing
a) Atmospheric pressure: 80kPa~106kpa
Storage and Transportation Environment Conditions
(1) Temperature: -20℃~55℃
(2) Relative humidity: 10% ~90%, non-condensing
(3) Atmospheric pressure: 70kPa~106kPa
Note: The storage and transportation environment conditions are the same whether the product is with original packaging or unpackaged.
Main Safety Standards
IEC 60601-1:2005 + AMD1:2012 + AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-2-57:2019 Medical electrical equipment - Part 2-57: Particular requirements for basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and aesthetic/cosmetic use
IEC 60601-1-2:2014 + AMD1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests