Semi-Automatic Blood Component Extractor
Times
2025年06月20日 15:19
Semi-Automatic Blood Component Extractor

点赞
投币
Instructions:
1. Plug in the power supply, turn on the power switch on the back of the device, the light at the reset button is on, pull the handle into the hook, and prepare for pulp separation.
2. Hang the whole blood bag into the hook at the front of the device, and snap the catheter into the sensor and the caliper.
3. Pull the handle out of the hook. After the splint completely presses the whole blood bag, shake the easy-break part left and right. After the easy-break part is completely broken, the separation will start automatically.
4. At the end of the plasma separation, the red blood cells trigger the sensor, the caliper is clamped, the separation of the plasma is stopped, the alarm sounds and the alarm light is on. The alarm sound will stop after 5 seconds, and the alarm light will stay on.
5. Close the clamp of the plasma bag manually, pull the handle into the hook, press the device reset button, the caliper will automatically reset, and the alarm will be released.
6. Break off the easy-break part of the preservation fluid bag, and the preservation solution flows into the whole blood bag. After the preservation solution has run out, remove the separated red blood cells and plasma bags, and the separation is completed.